implant placement validation
The maximum insertion torque for NobelActive™ was established at 70 Ncm, based on clinical study results(1):
• Implants were placed with varying insertion torque levels
• No correlation was found between complications and insertion torque at these levels, when compared (Unpublished data)
torque strength
NobelActive™ was designed with high torque strengths in mind:
• 282 Ncm maximum implant torque strength for NobelActive™ Ø3.5
• 452 Ncm maximum implant torque strength for NobelActive™ Ø4.3
When compared to the maximum insertion torque, a significant safety margin exists (see figure below). For further NobelActive™ mechanical testing information, see the NobelActive™ Technical Story.
Scientific evidence
1. Evaluation of NobelActive™ Implants. Five-year randomized controlled prospective multicenter study in 12 centers. Clinical Research Department, Nobel Biocare Services AG.